Yokogawa AQ6370C Optical Spectrum Analyzer OSA Repair and Calibration Services

Simbol Test Systems provides professional in-lab Yokogawa AQ6370C Optical Spectrum Analyzer repair and calibration services to help maintain accurate, reliable performance for your optical test and measurement equipment. Our experience with Ando and Yokogawa OSA platforms started in 2000 through technical collaboration with Ando, now part of Yokogawa.

For the AQ6370C OSA, Simbol has developed custom software to support automated calibration testing and update results directly in the instrument calibration tables where applicable. This process is designed to provide a more complete verification than a limited basic report and helps confirm the real operating condition of your optical spectrum analyzer.

Following calibration, your equipment may be returned with a calibration label, calibration certificate, and related calibration data. If the unit does not successfully complete calibration, we can evaluate the repair requirements and provide a repair quotation for approval before proceeding.

Traceability: Instrumentation used during this calibration is traceable to N.I.S.T. (National Institute of Standards and Technology) or C.N.R.C. (Canadian National Research Council).

Contact us with your model number and a brief description of the issue to request a calibration quote or repair evaluation for your Yokogawa AQ6370C Optical Spectrum Analyzer.

Yokogawa AQ6370C Optical Spectrum Analyzer Specification Sheet PDF Download Yokogawa AQ6370C Optical Spectrum Analyzer Specification Sheet (PDF)

Yokogawa AQ6370C Optical Spectrum Analyzer User Manual PDF Download Yokogawa AQ6370C Optical Spectrum Analyzer User Manual (PDF)

Read moreShow less
Reference:
AQ6370C_SER
Brand:
Add to compare0
Description

Yokogawa AQ6370C Optical Spectrum Analyzer – Repair and Calibration Services

The Yokogawa AQ6370C Optical Spectrum Analyzer is a high‑performance instrument widely used in telecom, R&D, and production environments. As these units age, proper calibration and maintenance become essential to ensure wavelength accuracy, power stability, and long‑term reliability. Simbol provides complete calibration and repair services for the AQ6370C, following strict OEM specifications and industry‑standard procedures.

Premium AQ6370C Calibration Services

Every AQ6370C sent to Simbol undergoes a full, traceable calibration designed to ensure the instrument meets or exceeds the manufacturer’s published specifications. Your unit will be returned with a calibration sticker, calibration certificate, and a complete set of before‑and‑after calibration data (8 pages of measurement results).

Calibration Items Included

  • Software Upgrade
  • Parameter & Data Memory Clear
  • Optical Alignment
  • Wavelength Calibration with Internal Cell
  • Wavelength Accuracy
  • Wavelength Linearity
  • Wavelength Resolution Accuracy
  • Wavelength Reproducibility
  • Level Flatness
  • Level Accuracy
  • Level Linearity
  • Polarization Dependency
  • Dynamic Range
  • Measurement Level Range
  • Maximum Input Level
  • Level Response (White Light Source)
  • Waveform Symmetry, Flatness and Ripple (DFB)
  • Gain (ASE)
  • Wide Band Level Stability (Broadband Light Source)

Traceability

All calibration work is performed using instrumentation traceable to N.I.S.T. (National Institute of Standards and Technology) or C.N.R.C. (Canadian National Research Council), ensuring full measurement traceability and compliance.

AQ6370C Repair and Mechanical Adjustment

If your AQ6370C does not pass calibration, Simbol provides complete repair services. We are the only independent laboratory in North America with the capability to mechanically realign Ando/Yokogawa monochromators, a critical procedure for restoring wavelength accuracy and optical performance.

After repairs are completed, your OSA is fully recalibrated to ensure it meets OEM performance standards and is returned in verified working condition.

Why Choose Simbol?

  • Specialized expertise with Yokogawa OSAs
  • Premium, traceable calibration procedures
  • Unique capability to realign monochromators
  • Fast turnaround and detailed documentation
  • Reliable repair capability for aging instruments

Maintain the accuracy and performance of your Yokogawa AQ6370C Optical Spectrum Analyzer with Simbol’s trusted calibration and repair services. Contact us today for a diagnostic evaluation or service quote.